How Often Should a Gas Detector Be Calibrated? A UK Guide
If you manage portable gas detectors, one question comes up again and again: how often should a gas detector be calibrated? The honest answer is "as often as the manufacturer and your risk assessment say", but in practice there is a well-established pattern most UK sites follow. This guide explains gas detector calibration intervals, what can shorten them, and how to keep on top of it without your monitors spending weeks out of service.
What gas detector calibration actually does
Every sensor in a gas detector drifts over time. Electrochemical sensors (used for gases such as CO and H2S) gradually lose sensitivity as they age, catalytic and infrared LEL sensors can be affected by poisons and contamination, and oxygen sensors are consumed as they work. Calibration exposes each sensor to a certified concentration of test gas and adjusts the instrument so that its reading matches the known value.
Without regular calibration, a monitor can still switch on, pass its self-test and display a reassuring zero while under-reading a real hazard. That is why calibration is not a paperwork exercise: it is the only way to confirm the readings are accurate.
How often should you calibrate a gas detector?
Most manufacturers recommend that portable gas detectors are calibrated at least every six months. Some instruments and some workplaces call for more frequent calibration, and a few manufacturers specify different intervals for particular models, so always check the user manual for your detector.
Alongside periodic calibration, most manufacturers recommend a bump test before each day's use. A bump test is a quick functional check that confirms the sensors respond to gas and the alarms work. It does not replace calibration, but it does catch a failed sensor between calibrations. We explain the distinction in more detail in our article on bump test vs calibration.
When to calibrate more often
Your own risk assessment may justify a shorter interval than six months. Common reasons include:
The monitor fails a bump test, or readings look unusual
The detector has been exposed to a high concentration of gas, or to known sensor poisons such as silicones
The instrument has been dropped, submerged or otherwise damaged
It is used heavily, in harsh, dusty or wet conditions, or in extreme temperatures
A sensor has been replaced or the instrument has been repaired
When the interval does not change
Leaving a detector in a drawer does not stop the clock. Sensors age whether or not the monitor is in use, so a detector that has sat unused for months still needs calibrating before it goes back into service.
What UK regulations expect
UK legislation does not set a single calibration interval for every gas detector. Instead, employers are expected to control risks properly, and where gas detection forms part of those controls the equipment needs to be suitable, maintained and working. In confined spaces, for example, the Confined Spaces Regulations 1997 require a risk assessment, and where atmospheric monitoring is part of the safe system of work, the monitor must be reliable.
In practice, following the manufacturer's recommended calibration interval and keeping a dated calibration certificate for each instrument is the clearest way to show your detectors are properly maintained.
Keeping calibration records
For every detector, keep a record of the serial number, the date of the last calibration, the next due date and the certificate. Bump test results are also worth logging, either on the instrument's own datalog or on a docking station. Every instrument we calibrate is returned with a calibration certificate, which makes audits much simpler.
How 7 Solutions handles gas detector calibration
Our manufacturer-trained engineers calibrate and service portable gas detectors from Honeywell (BW and RAE), Crowcon, Dräger, MSA, Industrial Scientific, WatchGas, GMI, Oldham and more. You can see the full list in gas detector brands we calibrate and repair.
Calibration prices (plus VAT) are:
Single gas monitors: £40
Multi-gas monitors: £90
PID monitors: £140
You can post instruments to our Hampshire service centre in Waltham Chase, Southampton, or arrange an on-site visit. Our Northern Service Centre covers the North East and Yorkshire. If you cannot be without detection while your monitors are away, our gas detector hire service can cover the gap.
Frequently asked questions
Is six-monthly calibration a legal requirement?
There is no single legal interval for all detectors. Six months is the interval most manufacturers recommend as a minimum, and following the manufacturer's guidance is the usual way to show equipment is properly maintained.
Can I calibrate my own gas detector?
It is possible with the right certified gas, regulator, software and training, but many sites prefer an independent service so that the work is done consistently and a certificate is issued for every instrument.
Does a bump test reset the calibration due date?
No. A bump test confirms the sensors respond, but it does not adjust them. The calibration due date only changes when a full calibration is carried out.
What if my detector fails calibration?
The sensor may need replacing or the instrument may need repair. We will advise you before carrying out any extra work.