Gas Detection for Farming and Slurry: A Practical UK Guide

Writer: Alex Graft
Alex Graft
11 minutes ago
5 min read

Slurry stores, reception pits, livestock buildings and silos are a routine part of farm work, and each can produce gases that give little or no warning. Hydrogen sulphide released when slurry is mixed is the best known, but ammonia, carbon dioxide, methane and, after silos are filled, nitrogen dioxide all need thought. This guide covers the gas hazards on farms, which gas monitor suits each job, and how to use and look after it.

The gas hazards in farming and slurry handling

A farm risk assessment will typically look at:

  • Hydrogen sulphide (H2S) from slurry. Stored slurry holds H2S, and agitating or mixing it can release a large amount very quickly. H2S is toxic and heavier than air, and at higher concentrations it can deaden the sense of smell, so the rotten-egg smell is not a reliable warning.

  • Ammonia (NH3). Given off by manure, urine and slurry, especially in housed livestock buildings. It is lighter than air and irritates the eyes and airways.

  • Carbon dioxide (CO2). Released from slurry and building up in poorly ventilated buildings. It is heavier than air and collects in pits, tanks and low areas, where it can displace oxygen.

  • Methane (CH4). A flammable gas produced by slurry, which can build up in covered stores and under slatted floors. Its lower explosive limit (LEL) is 4.4% by volume.

  • Nitrogen dioxide (NO2). Silo gas, produced by fresh silage in the days and weeks after a silo or clamp is filled. It is heavier than air and can gather in and around tower silos and feed rooms.

  • Oxygen depletion in slurry tanks, reception pits, silos and grain stores.

The UK workplace exposure limits (WELs) in HSE's EH40/2005 (fourth edition, January 2020) for the main farm gases are, as 8-hour TWA / 15-minute STEL: H2S 5 ppm / 10 ppm; ammonia 25 ppm / 35 ppm; carbon dioxide 5000 ppm / 15000 ppm; nitrogen dioxide 0.5 ppm / 1 ppm. WELs are legally binding where they apply under COSHH, but they are only one part of COSHH control, alongside ventilation, safe systems of work and keeping people out of danger areas. Our EH40 exposure limits reference lists the values for other common gases.

Which gas monitor fits which job

  • Single gas H2S monitor. For anyone mixing, pumping or spreading slurry, or working in livestock buildings. A personal H2S monitor is small, clips to the collar or chest and runs with little attention.

  • 4-gas monitor. Oxygen, flammable (LEL), carbon monoxide and H2S in one instrument, for work in or near confined spaces such as slurry tanks, reception pits, under-slat stores and silos. A pumped 4-gas monitor with a sample line lets you test the atmosphere from outside before anyone goes in. Note that a standard 4-gas monitor does not measure ammonia or carbon dioxide.

  • Multigas monitor. Five or six sensors, adding ammonia, carbon dioxide or nitrogen dioxide where the risk assessment calls for them, for example in livestock housing or around silos.

  • Fixed gas detection. For enclosed reception pits, under-floor stores or plant areas where gas can build up when nobody is there to carry a portable monitor. 7 Solutions supplies fixed systems to your specification and services them.

Slurry mixing and confined spaces

Slurry tanks, pits and silos are confined spaces, and the Confined Spaces Regulations 1997 require you to avoid entry where you reasonably can. If entry is unavoidable it needs a safe system of work, pre-entry gas testing and emergency arrangements in place before anyone goes in. Because H2S and CO2 sink while methane and ammonia rise, test at several levels, from the top of the space down, and keep monitoring continuously while work goes on. Where methane can build up, DSEAR also requires the fire and explosion risk to be assessed.

When mixing slurry, follow your own procedure: open the building up for ventilation, keep people and, where possible, livestock out, work from outside and upwind where you can, and do not go back in until your procedure says it is safe. A gas monitor is a warning device. It does not make an unsafe space safe.

Using a gas monitor day to day

  • Start-up. Switch on in clean air, let the monitor complete its self-test, and check the battery, the calibration due date and that the alarms sound.

  • Bump test. Many manufacturers recommend a bump test before each day's use: a short exposure to a known gas to confirm the sensors and alarms respond.

  • Wearing position. Wear it in the breathing zone, on the collar or chest, not under a coat, in a pocket or left in the tractor cab.

  • Alarm settings. EH40 does not set alarm levels. Settings follow your risk assessment, site standard, the purpose of the monitoring and what the instrument can do. An instantaneous alarm does not measure an 8-hour or 15-minute average, so an alarm set at a WEL does not by itself show compliance. Typical settings, depending on the site standard and risk assessment, are oxygen 19.5% low and 23.5% high, and flammable gas 10% and 20% LEL.

  • On alarm. Stop work, move out into fresh air, upwind if you can, warn others and do not go back until the cause is found. Never go into a pit, tank or building to help someone who has collapsed unless you are trained and equipped with breathing apparatus; call 999.

Calibration and record keeping

Calibration intervals vary by instrument and use, so follow the manufacturer; six months is common. Farm conditions are hard on monitors, and one that has been dropped, soaked, clogged with dust or exposed to a high reading should be checked before it is used again. Keep simple records of bump tests, calibration certificates and alarm events, so you can show the monitor was working when it was needed.

Our engineers calibrate and repair Honeywell, WatchGas, Crowcon, Dräger, MSA, Industrial Scientific, GMI, Oldham and more, with traceable calibration. Calibration and servicing for any brand costs £40 for a single gas monitor, £50 for a 4-gas and £90 for a multigas, ex VAT.

Slurry work is seasonal, so gas detector hire often makes sense for spreading season or a one-off tank clean: £35 a week for a single gas monitor, £45 for a 4-gas and £85 for a multigas, ex VAT. We also run gas detection training at £750 a day on site anywhere in the UK, or £350 for a 3-hour online session.

Frequently asked questions

Can I rely on smell to detect hydrogen sulphide?

No. At higher concentrations H2S can deaden the sense of smell, so the smell can fade just as the gas becomes more dangerous. A calibrated H2S monitor worn in the breathing zone gives a far more reliable warning.

Does a 4-gas monitor detect ammonia or carbon dioxide?

Not as standard. A typical 4-gas monitor measures oxygen, flammable gas, carbon monoxide and H2S. If ammonia or CO2 is a risk, you need a multigas monitor fitted with those sensors or a dedicated single gas monitor.

How often should a farm gas monitor be calibrated?

Follow the manufacturer's recommendation; six months is common. Many manufacturers also recommend a bump test before each day's use, and a monitor that has been damaged or exposed to high gas levels should be checked sooner.

Should I buy or hire a gas monitor for slurry work?

If you handle slurry regularly, owning a personal H2S monitor makes sense. For occasional jobs such as tank entry or cleaning, hiring a calibrated 4-gas or multigas monitor by the week is often simpler.

Talk to 7 Solutions

Not sure which monitor suits your farm, or need yours calibrated before spreading season? Call 7 Solutions on 01489 326031 or email sales@7solutions.co.uk and we will help you choose, hire or service the right gas detector.

 
 
 